This kind of meter increases a set of upper limit and lower limit electrical contact device on the basis of ordinary pressure gauge, when the equipment reaches the control pressure limit and sends a switching signal.
From the appearance, there are three Pointers, green Pointers, black Pointers, red Pointers. The green pointer is usually used for the lower limit control pointer, the black pointer is the instrument indicator pointer, and the red pointer is the upper limit control pointer. The upper and lower limits fluctuate through a knob on the outside of the glass cover to set the pressure range.
Next to the dial there is an electrical junction box, usually three lead terminals. In fact, the mutual relationship between the wiring is easy to judge, with a multimeter buzz gear can be simple measurement. After detection, one terminal is a common terminal, and the other two are upper and lower leads. (A moving pointer below the green pointer is normally closed, and a moving pointer above the red pointer is also normally closed.) .
Pressure gauge calibration table, detection is very convenient. Set the table and set the upper and lower limits as you like. For example, 0.2 ~ 0.8Mpa. Set the green pointer to 0.2mpa and the red pointer to 0.8mpa. Because the pressure is zero at this time, there must be a normally closed contact at the first detection, that is, the lower limit control contact, and then slowly pressurized over 0.2mpa, normally closed to normally open. Then continue to pressure to 0.8mpa, at this time there is only one normally closed contact, that is the upper limit control contact. Therefore, the relationship between their three terminals is easy to judge.
With our household switch is not different, one is to use manual, one is through the pressure to make the end of the spring tube displacement, through the gear transmission mechanism amplification, to drive the electrical contact device of the mobile contact action, so as to achieve the contact breaking and closing.