what is the reason for dropped needle of the pressure gauge

Jul 20, 2021

The reason for the dropped needle of pressure gauge 


33

There are two situations in which the "needle drop" of the pressure gauge is divided into two situations. One is that the needle is found to be dropped before the instrument is used after the instrument arrives; the other is that the indicator pointer falls off during use of the instrument.

      In either case, the needle drop problem of the pressure gauge should be one of the most serious failures of the instrument. It is thought that once the indicator pointer falls off, it will lose the pressure indication function, that is, the pressure gauge loses the most basic use function. This is Not acceptable to any user.

      The reasons for the needle drop of the meter after the meter arrives or before use generally have the following two aspects:

      1. After the instrument is ridiculed, there is no stitching or the stitching is uneven;

      2. Improper handling of the instrument during transportation, large vibration will cause the pointer to shake off the stainless steel pressure gauge;

      The reasons why the pointer falls off during the use of the meter are generally as follows:

      1. After the pointer is adjusted, the needle is not nailed or the nail is not strong;

      2. The taper of the inner hole of the needle library is inconsistent with the taper of the central axis of the movement, causing the nail to be weak;

      3. The needle library and the needle plate are not connected firmly, and the needle plate and the needle library are separated due to vibration or impact;

      4. The pressure of the meter is released too fast, and the pointer repeatedly hits the plate stop nail violently, causing the pointer to fall off;

      5. The working environment of the instrument has large amplitude and high frequency, and long-time work causes the pointer to fall off the precision pressure gauge.

      If the needle drop problem occurs in the meter, whether it occurs before use or during use, and regardless of the cause, the primary responsibility for the meter lies with the manufacturer. Because as long as the needle is firmly nailed when the meter is adjusted, the needle drop problem should not occur in the case of back concentration.