What are the use and maintenance methods of electric contact bimetal thermometer?
Use and maintenance methods of electric contact bimetal thermometer:
1. The length of the bimetallic thermometer protective tube of the electric contact inserted into the measured medium must be greater than the length of the temperature sensing element (generally, the insertion length should be greater than 100mm, and the insertion length of 0~50 ℃ range should be greater than 150mm), so as to ensure the accuracy of the measured value.
2. During the use and installation of electric contact bimetal thermometer, collision with the protective tube shall be avoided, and the protective tube shall not be bent or deformed.
3. Before use, check whether the electric contact bimetal thermometer exceeds the guarantee period (18 months). If it is expired, it can be used only after it has passed the re inspection.
4. The temperature value of electric contact bimetal thermometer for regular operation shall be 1/2~3/4 of the maximum range. During installation, the hexagon part shall be clamped to tighten the thread. It is strictly prohibited to tighten the thread by rotating the electric contact bimetal thermometer head.
5. When the electric contact bimetal thermometer measures or controls the medium temperature above 200 ℃, in addition to installing the joint to ensure sealing, attention should also be paid to the impact of radiation on the electric contact bimetal thermometer.
6. In order to increase the power of the electric contact and prevent the electric contact from burning out, the sensitive relay shall be used in the control circuit. It is recommended to use electronic relay with self-locking function to increase the power of electric contact.